SL 80 C
Specification:SL 80 C
Product:TURBOVAC SL 80 C Part No.:800002V3008 | |
Inlet connection: | DN 63 ISO-K |
Outlet connection: | DN 16 ISO-KF |
Pumping speed | |
N2 - Nitrogen: | 70 l/s |
Ar - Argon: | 65 l/s |
He - Helium: | 50 l/s |
H2 - Hydrogen: | 45 l/s |
Gas throughput | |
N2 - Nitrogen: | 3.5 mbar x l/s |
Ar - Argon: | 3.5 mbar x l/s |
He - Helium: | 2.0 mbar x l/s |
H2 - Hydrogen: | 1.0 mbar x l/s |
Compession ratio | |
N2 - Nitrogen: | > 2.0 x 106 |
Ar - Argon: | > 2.0 x 106 |
He - Helium: | 5.0 x 102 |
H2 - Hydrogen: | 2.0 x 102 |
Ultimate pressure: | < 5.0 x 10-8 mbar < 3.8 x 10-8Torr |
Max. foreline pressure for N2 : | 3.5 x 10-1 mbar 2.6 x 10-1 mbar |
Nominal rotation speed: | 72000 min-1 72000 rpm |
Run -up time: | ≈ 1.5 min |
Max. power consumption: | 120 W |
Power consumption at ultimate pressure: | 17 W |
Protection rating: | IP 20 |
Admissible ambient temperature: | +15 to +45°C +59 to +113°F |
Cooling standard: | Convection |
Cooling optional: | Air / Water |
Cooling water connection: | G 1/8", Inside thread / 8 mm hose nozzle |
Cooling water consumption: | 15 - 60 l/h |
Permissible cooling water pressure: | 2 - 7 bar |
Permissible cooling water temperature: | +10 to +40°C +50 to +104°F |
Dimensions: | see dimension sheet |
Weight: | ≈ 1.9 kg ≈ 4.2 lbs |